Agilent 8453 UV Vis spectrophotometer G1103A
Spectrophotometers
Agilent
- EQUIPMENT DESCRIPTION
- SPECIFICATIONS
- INCLUDED
- NOTES
The Agilent UV Vis 8453 is a spectrophotometer that delivers fast results and repeatability with low noise and stable, high throughput optics. This compact UV vis spectrophotometer uses a diode-array and prealigned deuterium and tungsten lamp light sources for fast spectral scanning across the ultra-violet to visible light spectrum to return spectral acquisition and analysis in less than one second. This UV-Vis spectrophotometer has a wavelength before 190-1100 nm, and a 1nm slit width, cutting down on stray light to < 0.03 %.
The Agilent UV Vis 8453 spectrophotometer G1103A supports compliance with GLP regulations, with its own clock for time and date stamps and self-test procedures to make sure performance is consistent in between validations. This spectrophotometer has an open sample area for easy loading of samples and makes large accessories easy to use. This system is ideal for labs needing to conform to GLP regulations, or for more generalized tasks. Users of this uv-vis spectrophotometer are guided through a graphical interface, that provides at a glance confirmation of actions and statuses, and quick access to frequently used programs.
Agilent 8453 UV-Vis Spectrophotometer Specifications:
Wavelength range | 190–1100 nm |
Slit width | 1 nm |
EP resolution test | > 1.6 toluene in hexane, ratio abs. at 269 nm/266 nm |
Stray light | < 0.03 % at 340 nm (NaNO2, ASTM) < 0.05 % at 220 nm (NaI, ASTM) < 1 % at 198 nm (KCl, EP) |
Wavelength accuracy | < ± 0.5 nm 0.5-second scan (NIST 2034) < ± 0.2 nm at 486.0 and 656.1 nm |
Wavelength reproducibility | < ± 0.02 nm ten consecutive scans (NIST 2034) |
Photometric accuracy | < ± 0.005 A at 440.0, 465.0, 546.1, 590.0, and 635.0 nm, 1 A (NIST 930e) < ± 0.01 A at 235, 257, 313, 350, 430 nm, (potassium dichromate, EP method) |
Photometric noise | < 0.0002 A sixty 0.5 second scans at 0 A, 500 nm, rms |
Photometric noise | < 0.0002 A sixty 0.5 second scans at 0 A, 500 nm, rms |
Photometric stability | < 0.001 A/h at 0 A, 340 nm, after 1-hour warm up, measured over 1 hour, every 5 seconds, constant ambient temp. |
Baseline flatness | < 0.001 A 0.5-second blank, 0.5-second scan, RMS |
Typical scan time | 1.5-second full range |
Shortest scan time | 0.1-second full range |
Time until next scan | 0.1-second full range, 0.1-second scan, up to 150 consecutive scans |
Dimensions | 24x16x8” |
Weight | 50lbs |
Power | 100-120/220-240V, 50/60 Hz |
- Agilent 8453 UV-Vis Spectrophotometer
- Windows 7 Notebook with UV-Vis Chemstation
- System Validation Report
- 120V Power Cord
- Communication Cable
- Agilent 8453 User Manual
- 30-Day Warranty
This Agilent 8453 UV-Vis Spectrophotometer has been fully inspected and comes with a current system validation report. All of our lab equipment comes backed by our standard warranty. Should you have any questions, please don't hesitate to contact us.