Eppendorf Biophotometer Spectrophotometer

Availability: Out of stock SKU: 16713 Category:

Description

Eppendorf Biophotometer Spectrophotometer

The Eppendorf Biophotometer 6131 Spectrophotometer is a compact, easy-to-transport cuvette photometer that’s fast, reliable, and simple to use. This single-beam photometer is ideal for routine work in molecular biology, biochemistry, or cell biology. The Biophotometer uses 6 wavelengths and 12 preprogrammed methods and is optimized for small sample volumes so you keep more of your precious samples. Methods for this Eppendorf photometer include nucleic acid calculations for double and single-strand DNA, Oligo, proteins methods including Bradford and Lowry, as well as bacteria density.

Users of this Eppendorf Biophotometer can use standard rectangular glass or plastic cuvette such as the disposable UVette. When not in use a cover protects the shaft from dust and other contaminants.

Measurements are performed with the push of a button, with fixed keys for quick access to groups of programs. The BioPhotometer plus converts the measured absorbance values into concentration results, and displays it on screen or exported via a printer (sold separately). Data can be transferred to a PC along with all the calibration data for printing, or further processing and archiving.

Specifications

Eppendorf BioPhotometer Spectrophotometer Specifications:

Measuring principle Absorption single beam photometer with reference
Light source Xenon flash lamp
Monochromator Holographic concave grating
Beam receiver Silicon photodiodes
Measuring wavelengths 230, 260, 280, 320, 562, 595 nm
Wavelength selection Method-dependent, program-controlled
Spectral bandwidth 5 nm at 230 to 320 nm

7 nm at 562 to 595 nm

Wavelength inaccuracy ± 1 nm at 230 to 280 nm

± 2 nm at 320 to 595 nm

Photometric measuring range Quartz glass cuvette 0.000 to 3.000 A

UVette – 2.5A at 230nm

2.6A at 260nm

2.8A at 280 nm

2.9A at 320 nm

Photometric Imprecision ≤0.002A at 0A

≤ 0.005A at 1A

Photometric inaccuracy ± 1% at 1 A
Accuracy of reading 0.001 A
Stray-light component < 0.05%
Cuvette material For dsDNA, ssDNA, RNA, Oligo, Protein: Quartz glass or plastic (UVette by Eppendorf)

For OD600, Bradford, Lowry, BCA: Glass or plastic

Cuvette shaft 12.5 mm x 12,5 mm, not temperature-controlled
Height of the light beam in the cuvette 8.5mm
Light beam in the cuvette Width: 1 mm

Height: 1.5 mm

Memory 12 preprogrammed, modifiable method programs
Power 100 to 240 V ± 10%: 50 to 60 Hz
Dimensions 8x13x3”
Weight 13 lbs

Included

Included with the Eppendorf Biophotometer

Eppendorf Biophotometer | CAT# 6131-20522
Thermal Printer
120V Power Cord
30-Day Warranty

Notes

Technical Notes

This photometer has been fully inspected and is ready to ship. All of our lab equipment comes backed by our standard warranty. Should you have any questions, please don’t hesitate to contact us.

Video

Eppendorf Biophotometer Video

Eppendorf Biophotometer 6131 ID 16713